Prompt
Calibration
For security reasons, the device only by suitably qualified personnel operation and maintenance. Prior to the operation or maintenance, please read and fully manage all solutions to these instructions. Including operations, maintenance of equipment and process methods. And a very important safety precautions. Read the following Cautions before using the detector. Table 1 CautionsCautions |
1. Warning: Unauthorized replacement of replacement parts in order to avoid the impact of the instrument Normal use. 2. Warning: Do not disassemble, heat or incinerate batteries. Otherwise battery possible explosion, fire or chemical burn hazard. 3. Warning: Do not calibrate the instrument in hazardous locations or set parameters. 4. Warning: all factory pre-calibrated instrument. Users use recommended calibration once at least six months in order to maintain a quasi-instrument Accuracy. 5. WARNING: Be sure to avoid using the instrument in corrosive atmospheres. 6. Warning: Do not use solvents, soaps, cleaning or polishing agents outside Shell. |
Maintenance
To maintain the detector in good operating condition, perform the following basic maintenance as required: • Calibrate, bump test, and inspect the detector at regular intervals. • Maintain an operations log of all maintenance, calibrations, bump tests, and alarm events. • Clean the exterior with a soft damp cloth. Do not use solvents, soaps, or polishes. • Do not immerse the detector in liquids. Table 10 Replacing the BatteryFigure 1
Appearance description as shown in Table 2 Table 2 Item | Description |
1 | Sensor |
2 | Buzzer (audible alarm) |
3 | Pushbuttons |
4 | Mask |
5 | Liquid crystal display(LCD) |
6 | Visual alarms bars(LEDs) |
7 | Alligator clip |
8 | Nameplate |
9 | Product ID |
Calibration gas detection alarm connection diagram shown in Figure 3, Table 8 for calibration connection diagram shows.
Figure 3 Connection diagram
Figure 2 Display Elements
Table 3 Display Elements DescriptionItem | Description |
1 | Numeric value |
2 | Battery(Display and flashes when the battery is low) |
3 | Parts per million (ppm) |
Measured gas | Gas Name | Technical Specifications | ||
Measuring range | Resolution | Alarm | ||
CO | Carbon monoxide | 0-1000ppm | 1ppm | 50ppm |
H2S | Hydrogen sulfide | 0-100ppm | 1ppm | 10ppm |
NH3 | Ammonia | 0-200ppm | 1ppm | 35ppm |
PH3 | Phosphine | 0-1000ppm | 1ppm | 10ppm |
Item | Function |
Standby mode, the menu button | |
Long press for power on and off button | |
Note: | |
1. To start gas detection alarm, press and hold the button for 5 seconds. After the gas detection alarm through a self-test, then begin normal operation. | |
2. To turn off the gas detection alarm, press and hold the button for 5 seconds. | |
Menu operation is on the turn, button backlight switch | |
Shift buttons for menu operation | |
Menu operation is OK function, clear the alarm button |
Display | Description |
ALA1 | Setting low alarm |
ALA2 | Setting high alarm |
ZErO | Cleared (operating in the pure air) |
-rFS. | Restore the factory default password 2222 |
Table 8 Part Description | |
Item | Description |
① | Gas Detector |
② | Calibration cap |
③ | Hose |
④ | Regulator and gas cylinder |
Procedure | Screen |
Hold down the button and press the button, release | 1100 |
Enter the 1111 switch and flashing bit by and | 1111 |
Press the button | idLE |
Double-click the button | 2-FU |
Press the button, Will display the default calibration gas concentration value | 0500 (calibration gas concentration value) |
The actual value of the input switching concentration calibration gas flashing and blinking bit by bit on the key and keys. | 0600 (eg) |
Press the button, Screen '-' move from left to right. After displaying good, then display idLE. | idLE |
Long press the button,return to the concentration detection interface , such as the calibration is successful, the concentration of the calibration value will be displayed, if the difference between the value of the standard gas concentration is large, the above operation again. | 600 (eg) |
Item | Description | Detector parts diagram |
① | Rear shell machine screws | |
② | Rear shell | |
③ | Battery | |
④ | PCB | |
⑤ | Sensor | |
⑥ | Front shell |
Questions and answers
1. The measured value is not accurate The gas detection alarm after a period of time used to detect concentrations may occur deviation, periodic calibration. 2. Concentration exceeds the set alarm value; there is no sound, light or vibration alarm. Refer to Chapter 7 [Special instructions], the settings -AL5 inside to ON. 3. The battery inside the gas detection alarm can charge? You can not charge, replace the battery power is exhausted after. 4. Gas detection alarm can not boot a)Gas detection alarm crashes, open the detector housing, remove the battery, and then reinstall it. b)The battery runs out, open the detector housing, remove the battery, and replace the same brand, same model battery. 5. What is the fault code information? Err0 password error Err1 set value is not within the allowed range Err2 calibration failure